BRONKHORSTHIGH‐TECH
Page8 MASS‐VIEW®series 9.17.051
Thepressurelossacrosstheneedlecanbedescribedby:
∆P
k
T∅
P
120°1
The spring force in the upstream membrane valve closes the valve in the “zero” flow situation. By
openingtheneedlevalvetheΔPwillincrease,sofarthatthemembraneliftsupandcompensatesthe
springforce.Intheequilibriumpositiontheflowwillbeconstantevenwhentheupstream
pressure(P
in
)
changes.Howeverwhentheupstreampressureincreases,itwillcauseasmalleffectonthespringforce,
which means thattheΔP to open the membrane valve will be a littlesmaller. Thisresults in a lower
flow.
Inpractisethedeviationoftheidealstraightlinewillbe
lessthan0,5‐1%perbar.
1.11 Electronics, display and 4-way button
The MASS‐VIEW
®
has a bright graphical OLED display,
clearly visible over any angle, which allows reading of
actualflow(valueandbargraph),totalflowandgastype.
Thethree displayoptions can be changedusingthe4‐way
button This button is a navigation joystick which has
UP/DOWN/LEFT/RIGHT keys and a centre push (ENTER)
key.
4‐waybutton