Instruction manual Digital readout and control system E-7000 for analog and digital instruments Doc. no.: 9.17.004N Date: 25-07-2011 ATTENTION Please
BRONKHORST HIGH-TECH B.V. page 10 9.17.004 1.4.3.1 Analog input/output signals Analog input signals should be connected to pin 2 (+) and 0 V/comm
BRONKHORST HIGH-TECH B.V. 9.17.004 page 11 1.4.6 EMC and cables All system setups described in this manual carry the CE-mark. Therefore they hav
BRONKHORST HIGH-TECH B.V. page 12 9.17.004 1.5 Specifications 1.5.1 Housings 238.5 mm213.4 mm213.4 mm235.4 mm238.5 mm232.5 mm140 mm132.5 mm112
BRONKHORST HIGH-TECH B.V. 9.17.004 page 13 426.8 mm448.8 mm426.8 mm112.5 mm132.5 mm132.5 mm232.5 mm238.5 mm282.514 TE76 mm260 mm133.4
BRONKHORST HIGH-TECH B.V. page 14 9.17.004 1.5.2 EMC and housing All electronic circuits and modules have been designed to meet the requirements
BRONKHORST HIGH-TECH B.V. 9.17.004 page 15 2 INSTALLATION 2.1 General 2.1.1 Receipt of equipment Check the outside packing for damage incurred
BRONKHORST HIGH-TECH B.V. page 16 9.17.004
BRONKHORST HIGH-TECH B.V. 9.17.004 page 17 3 Operation of analog instruments analog instrument 3.1 Ge
BRONKHORST HIGH-TECH B.V. page 18 9.17.004 3.2 Startup and menu selection LoadingParameters ...(0000 0000)Bronkhorst Hi-Tec(0001 0000)Checking fl
BRONKHORST HIGH-TECH B.V. 9.17.004 page 19 Menu descriptions 0001 0000 Startup screen showing company name 0002 0000 Startup screen showing
BRONKHORST HIGH-TECH B.V.
BRONKHORST HIGH-TECH B.V. page 20 9.17.004 3.3 Measure menu EDIT SETPOINT:xxx.xx (step)(1100 0000)CHANGE SETPOINT:SETPOINT ZERO(1400
BRONKHORST HIGH-TECH B.V. 9.17.004 page 21 Menu descriptions 1100 0000 Stepsize editor for editing setpoint/slave-factor by a 0.1% step up or do
BRONKHORST HIGH-TECH B.V. page 22 9.17.004 3.4 Operation menu OPERATION MENU(2000 0000)CONTROL MODEFLOW-BUS(2140 0000)OPERATION MENUCONTROL MODE(2
BRONKHORST HIGH-TECH B.V. 9.17.004 page 23 Menu descriptions 2100 0000 Operation mode. Here you can change the source of the setpoint for the op
BRONKHORST HIGH-TECH B.V. page 24 9.17.004 3.5 Reset menu RESET MENU(3000 0000)RESET MENURESET ALARM(3100 0000)RESET MENUPASSWORD(3300 00
BRONKHORST HIGH-TECH B.V. 9.17.004 page 25 Menu descriptions 3100 0000 Reset Alarm. Here you can reset an alarm if it is active. This will resu
BRONKHORST HIGH-TECH B.V. page 26 9.17.004 3.6 Counter menu COUNTER MENU(4000 0000)COUNTER MENURESET(4100 0000)COUNTER MENUMODE(4200 0000)COUNTER
BRONKHORST HIGH-TECH B.V. 9.17.004 page 27 Menu descriptions 4100 0000 Counter Reset Enable. Here you can enter how the counter may be reset.
BRONKHORST HIGH-TECH B.V. page 28 9.17.004 3.7 Alarm menu ALARM MENU(5000 0000)ALARM MENURESET(5100 0000)ALARM MENUMODE(5200 0000)ALARM MENULIMIT(
BRONKHORST HIGH-TECH B.V. 9.17.004 page 29 Menu descriptions 5100 0000 Alarm Reset Enable. Here you can enter which way the alarm may be rese
BRONKHORST HIGH-TECH B.V. Even though care has been taken in the preparation and publication of the contents of this manual, we do not assume leg
BRONKHORST HIGH-TECH B.V. page 30 9.17.004 3.8 Instrument menu INSTRUMENT MENU(6000 0000)INSTRUMENT MENUINSTRUMENT TYPE(6100 0000)INSTRUMENT ME
BRONKHORST HIGH-TECH B.V. 9.17.004 page 31 Menu descriptions 6100 0000 Instrument Type. Here you can change if the instrument will control or o
BRONKHORST HIGH-TECH B.V. page 32 9.17.004 3.9 Local menu LOCAL MENU(7000 0000)LOCAL MENUREADING MODE(7200 0000)LOCAL MENUDISPLAY MODE(7100 0000)L
BRONKHORST HIGH-TECH B.V. 9.17.004 page 33 Menu descriptions 7100 0000 Display Mode. Here you can change the values that will be displayed on
BRONKHORST HIGH-TECH B.V. page 34 9.17.004 3.10 FLOW-BUS menu FLOW-BUS MENU(8000 0000)FLOW-BUS MENUMODE(8100 0000)FLOW-BUS MENUSETTING(8200 0
BRONKHORST HIGH-TECH B.V. 9.17.004 page 35 Menu descriptions 8100 0000 FLOW-BUS Mode. Here you can install this module on the FLOW-BUS or star
BRONKHORST HIGH-TECH B.V. page 36 9.17.004 3.11 Calibration with polynomial functions 3.11.1 General information A normally calibrated device w
BRONKHORST HIGH-TECH B.V. 9.17.004 page 37 3.11.4 Presentation of parameters Parameters 'a - d' and 'e - h' are polynomial fu
BRONKHORST HIGH-TECH B.V. page 38 9.17.004
BRONKHORST HIGH-TECH B.V. 9.17.004 page 39 FLOW-BUS 4 Operation of digital (FLOW-BUS) instruments di
BRONKHORST HIGH-TECH B.V.
BRONKHORST HIGH-TECH B.V. page 40 9.17.004 4.2 Startup and menu selection LoadingParameters...(0000 0000)BRONKHORSTHI-TEC(0001 0000)Checking flowb
BRONKHORST HIGH-TECH B.V. 9.17.004 page 41 Menu descriptions 0000 0000 Startup screen showing message loading parameters. 0001 0000 Startup scr
BRONKHORST HIGH-TECH B.V. page 42 9.17.004 4.3 Measure menu EDIT SETPOINT:xxx.xx (step)(1100 0000)CHANGE SETPOINT:SETPOINT ZERO(1400 000
BRONKHORST HIGH-TECH B.V. 9.17.004 page 43 Menu descriptions 1100 0000 Stepsize editor for editing setpoint/slave-factor by a 0.1% step up or dow
BRONKHORST HIGH-TECH B.V. page 44 9.17.004 4.4 Operation menu OPERATION MENU(2000 0000)CONTROL MODEFLOW-BUS(2110 0000)OPERATION MENUCONTROL MODE(2
BRONKHORST HIGH-TECH B.V. 9.17.004 page 45 Menu descriptions 2100 0000 Operation mode. Here you can change the source of the setpoint for the ope
BRONKHORST HIGH-TECH B.V. page 46 9.17.004 4.5 Reset menu RESET MENU(3000 0000)RESET MENURESET ALARM(3100 0000)RESET MENUPASSWORD(3300
BRONKHORST HIGH-TECH B.V. 9.17.004 page 47 Menu descriptions 3100 0000 Reset Alarm. Here you can reset an alarm if it is active. This will resul
BRONKHORST HIGH-TECH B.V. page 48 9.17.004 4.6 Counter menu COUNTER MENU(4000 0000)COUNTER MENURESET(4100 0000)COUNTER MENUMODE(4200 0000)COUNTER
BRONKHORST HIGH-TECH B.V. 9.17.004 page 49 Menu descriptions 4100 0000 Counter Reset Enable. Here you can enter how the counter may be reset.
BRONKHORST HIGH-TECH B.V. TABLE OF CONTENTS 1 Introduction 1.1 General description ...
BRONKHORST HIGH-TECH B.V. page 50 9.17.004 4.7 Alarm menu ALARM MENU(5000 0000)ALARM MENURESET(5100 0000)ALARM MENUMODE(5200 0000)ALARM MENULIMIT(
BRONKHORST HIGH-TECH B.V. 9.17.004 page 51 Menu descriptions 5100 0000 Alarm Reset Enable. Here you can enter which way the alarm may be reset
BRONKHORST HIGH-TECH B.V. page 52 9.17.004 5700 0000 Alarm delay 5710 0000 Here you can enter the alarm delay in seconds. This delay is used fo
BRONKHORST HIGH-TECH B.V. 9.17.004 page 53
BRONKHORST HIGH-TECH B.V. page 54 9.17.004 4.8 Instrument menu INSTRUMENT MENU(6000 0000)INSTRUMENT MENUINSTRUMENT TYPE(6100 0000)INSTRUMENT MENUS
BRONKHORST HIGH-TECH B.V. 9.17.004 page 55 Menu descriptions 6100 0000 Instrument Type. Here you can change if the instrument will control or on
BRONKHORST HIGH-TECH B.V. page 56 9.17.004 6A00 0000 Restore Backup. This function restores the backup (the factory settings) which was made aft
BRONKHORST HIGH-TECH B.V. 9.17.004 page 57
BRONKHORST HIGH-TECH B.V. page 58 9.17.004 4.9 Local menu LOCAL MENU(7000 0000)LOCAL MENUREADING MODE(7200 0000)LOCAL MENUDISPLAY MODE(7100 0000)L
BRONKHORST HIGH-TECH B.V. 9.17.004 page 59 Menu descriptions 7100 0000 Display Mode. Here you can change the values that will be displayed on
BRONKHORST HIGH-TECH B.V. 3.11.3 Polynomial functions of sensor signal and setpoint...
BRONKHORST HIGH-TECH B.V. page 60 9.17.004 4.10 FLOW-BUS menu FLOW-BUS MENU(8000 0000)FLOW-BUS MENUINSTALL MYSELF(8100 0000)FLOW-BUS MENUI
BRONKHORST HIGH-TECH B.V. 9.17.004 page 61 Menu descriptions 8100 0000 Install myself. Here you can install this module on the FLOW-BUS and st
BRONKHORST HIGH-TECH B.V. page 62 9.17.004 NOTE 2: Make sure this module has been switched-off before connecting it to the FLOW-BUS. First make
BRONKHORST HIGH-TECH B.V. 9.17.004 page 63 4.11 Calibration with polynomial functions 4.11.1 General information A normally calibrated device will
BRONKHORST HIGH-TECH B.V. page 64 9.17.004 4.11.5 Using polynomial functions at readout/control unit The parameters for the polynomial functions
BRONKHORST HIGH-TECH B.V. 9.17.004 page 65 5 MAINTENANCE - Be sure you operate under the correct input specifications. - Be sure the syste
BRONKHORST HIGH-TECH B.V. page 66 9.17.004 6 TROUBLESHOOTING • LCD is not lit after 'power on' • check power input • check fuse •
BRONKHORST HIGH-TECH B.V. 9.17.004 page 7 1 INTRODUCTION 1.1 General description The Bronkhorst High-Tech B.V. serie E-7000 is a modular digital
BRONKHORST HIGH-TECH B.V. page 8 9.17.004 Code for external output/input signals: N N - A A A Sensor output signal Code 0 - 5 Vdc
BRONKHORST HIGH-TECH B.V. 9.17.004 page 9 for digital instruments (FLOW-BUS): 1.4.2 Connection to measuring and controlling devic