BRONKHORSTHIGHTECH
9.17.051 MASSVIEW®series Page11
2.3 Mounting
ThemountingpositionoftheMASSVIEW
®
isvertical,althoughitispossibletoturnthedevice180°and
change the display setting. Avoid installation in close proximity of mechanic vibration and/or heat
sources.
2.4 In-line filter
2.5 Fluidic connections
The MASSVIEW
®
can be equipped with (compression) fittings. Tighten the nut finger tight; while
holding the instrument, then tighten the nut 1 turn. If applicable follow the guidelines of the
manufacturer of the fittings. While tightening fittings, do not apply excessive force, in order to avoid
damagingin/outputthreadorothersensitive
partsofyourinstruments.

PleasebeawarewhenusingOxygen(O
2
)thatthefittingsmountedtothebodyshouldbeabsolutely
oxygenclean.
2.6 Piping
Besurethatpipingisabsolutelyclean!
DO NOT install small diameter piping on high flow rates, because the inlet jet flow will affect the
accuracy.
DONOTmountabruptanglesdirectonin‐andoutlet,especiallynotonhighflowrates.Atleast10pipe
diametersdistancebetweentheangleandtheinstrumentisrecommended.
DO
NOT mount pressure regulatorsdirect on the inlet of gas flowmeters/controllers,but allow some
metersofpiping(atleast25pipediameters).
2.7 Electrical connections
TheMASSVIEW
®
instrumentsarepoweredwith+15Vdcto+24Vdc.
When providing your own power supplybe surethat voltageandcurrent ratingare according to the
specifications of the instrument(s) and furthermore that the source is capable of delivering enough
powertotheinstrument.RefertoHookupMASSVIEW
®
series(document nr. 9.16.085) for more
details.
When using other cables, cable wire diameters should be sufficient to carry the supply current and
voltagelossesmustbekeptaslowaspossible.
Whenconnectingthesystemtootherdevices(e.g.toPLC),besurethattheintegrityoftheshieldingis
notaffected.Donotuseunshieldedwireterminals.
Althoughthegasestobemeasuredshouldbeabsolutelyfreeofdirt,oil,moistureandotherparticles,
itis recommendedtoinstallaninlinefilter upstreamoftheflowmeter/ controller,and if backflow
can occur, a downstream filter is recommended too. Be aware of the pressure drop
caused by the
filter.
Always check your system for leaks, before applying fluid pressure. Especially if toxic, explosive or
otherdangerousfluidsareused.
Werecommend usingourstandard cables.Thesecableshave the rightconnectorsandif looseends
areused,thesearecolouredandmarkedtopreventimproperconnection.
Allcablingoptionscanbepurchasedonlinewww.massflowonline.com