Mode(FlowDDEparameter156)
Withthe“mode”parametertheusercanchooseifthealarmsresetautomaticallyormanually.
Whentheresetmodeissettoautomatic,thealarmswillberesetwhenmeasurereachesthedesired
levelandthedelaytimehasexceeded.
Whentheresetmodeismanual,
theusercanresetthealarmsinthefollowingways:
1. Withthe“reset”optioninthealarmsmenu.
2. ViaRS‐232(orFLOWDDE/FlowView/FlowPlot).
Writingthevalue“2”toFLOWDDEparameter114resetsthealarms.
3. Whenthereadoutscreenisvisibleandanalarmis
active,thiswillbedisplayedatthetopof the
screen (see chapter 3.4.7). By pressing up at an alarm situation, the user can reset the alarms
fromthereadoutscreen.
Relay
The relays parameter sets the output mode of the relays. The user can choose between disabled,
enabledand
pulse.Whenarelayisactive,therelaycontactwillconnecttothe0Vsignal.
3.4.7 Alarm outputs
Thestatusofthealarmscanbereadinvariousways.
Forexamplewhentheminimumandmaximumalarmsareenabledthisis
madevisibleinthereadoutscreenwithtwomarkersattheleftofthe
graph(seepictureontheright).
Whenanalarmarisesthecorrespondingmarker
willbehighlightedand
theheaderwillblinkanddisplaythealarm.
TheMASS‐VIEW
®
metershavetworelayswhichindicateanalarmsituation.
Thestatusoftherelaysdependonthealarmsourceandstatusofthe
relayparameter(seetablebelow).
disable enable pulse
Relay1 Relay2 Relay1 Relay2 Relay1 Relay2
Minalarm 0 0 1 0 0/1/0/1/0/1… 0
Maxalarm 0 0 0 1 0 0/1/0/1/0/1…
Response(min) 0 0 1 1 0/1/0/1/0/1… 0/1/0/1/0/1…
Response(max) 0 0 1 1 0/1/0/1/0/1… 0/1/0/1/0/1…
Counter 0 0 1 1 0/1/0/1/0/1… 0/1/0/1/0/1…
3.4.8 User tag
TheUsertagcanbeeditedatthemenuiteminfo.
3.5 Other gases
TheMASS‐VIEW
®
seriesarecalibratedonAir.Allotherpre‐installedgasesarecalculatedusinga
theoreticalconversionmodelwhichdescribestheflowcurveforothergaseswhichwillintroduce
additionalinaccuracies.
However,astheaccuracyoftheconversionmodelalsodependsonviscosity,pressureandtemperature,
specialattentionshouldbetaken
forgasesintheheterogeneousgasstatewherespecificheat,density,
thermalconductivityandviscositycanvarytremendously.
3.6 Zeroing procedure
TheMASS‐VIEW
®
hasazeroprocedureforeachindividualgas.Thezerofunctionpermitstheuserto
zerotheinstrumentwhenneeded.Pleasemakesuretheinstrumentiszeroedonthedesiredgasatthe
desiredoperatingtemperatureandpressureconditions.Itisextremelyimportantthatallflowbe
blockedandthe
MASS‐VIEW
®
isinastatic(NOFLOW)condition.SelectfromthemainMenuOption‐>
Fluidset‐>Zero.Pauseafewseconds,thenpressanykeytocompletethezeroprocedure.