Bronkhorst High-Tech B.V.
IQ+FLOW 9.17.04528
Value zero will be indicated by a period of 1 sec off (0.5 sec off + 0,5 sec off).
When value zero is wanted, press switch shortly and release it again within 1 sec.
Before each action of flash-counting, the LEDs to be used for counting will flash in a high frequency.
(Pattern: 0.1 sec on, 0.1 sec off). As soon as the switch is pressed-down, this LED (or both LEDs) will be off and the
counting sequence will start.
Changing node address, baud rate or parity in 'Configuration Mode' or in normal mode
The procedure for changing the baud rate, node address or parity for a FLOW-BUS or Modbus configuration over RS485 is described
in Section 3.5.3 (single-channel versions only). In the list below the selectable bus parameters with corresponding values are
presented.
When changing any of the parameters below in normal running/operation mode the communication with the
instrument may be lost. After a restart the instrument is available for communication on the newly selected
parameters. When in doubt activate the 'Configuration Mode' (38k4 and RS232 FLOW-BUS (Propar)) using the
micro-switch.
Fieldbus 2 selection
This parameter sets the fieldbus type. Parameter value 0 = FLOW-BUS, 1 = Modbus RTU, 2 = Propar, 3 = Modbus ASCII
Fieldbus 2 address
Set the node address at any of the allowed values listed in the table in Section 3.5.3, e.g. '3'.
Fieldbus 2 baud rate
0xFC48...0xFC49/64585...64586
Set the baud rate at any of the allowed values listed in table in section Section 3.5.3, e.g. '19200'.
Fieldbus 2 medium
Select the communication medium. Parameter value 0 = RS232, 1 = RS485
Fieldbus 2 parity
Set the parity for Modbus communication. Parameter value 0 = None, 1 = Odd, 2 = Even.