Bronkhorst High-Tech B.V.
IQ+FLOW 25
9.17.045
Dynamic Display Factor
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Float
RW
0…1.0
56
117/1
0xF508…0xF509/62729…62730
This value should not be changed.
Static Display Factor
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Float
RW
0…1.0
57
117/2
0xF510…0xF511/62737…62738
This value should not be changed.
CycleTime
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char
R
0…255
52
114/12
0x0E4C/3661
Note: The unit of parameter CycleTime is 10 ms. Example: value 0.2 means 2 ms
4.1.6 Alarm / Status parameters
i
www
See document nr. 9.17.023: Operational instructions for digital instruments, chapter 6.
http://www.bronkhorst.com/files/downloads/
manuals_english/917023_operation_instructions_digital_instruments.pdf
4.1.7 Counter parameters
i
www
See document nr. 9.17.023: Operational instructions for digital instruments, chapter 7.
http://www.bronkhorst.com/files/downloads/
manuals_english/917023_operation_instructions_digital_instruments.pdf
4.2 Special instrument features
4.2.1 Auto zeroing
To start the auto zero-procedure by digital operation two parameters should be written:
Control Mode
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char
RW
0…255
12
1/4
0x0024/37
Calibration Mode
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char
RW
0…255
58
115/1
0x0E61/3682
Value
Mode
Instrument action
0
Idle
Idle
9
Auto zero
Auto-zeroing
255
Error
Idle