Doc. no.: 9.17.045Q Date: 14-04-2015Instruction ManualPlease read this Instruction Manual carefully before installing and operating the instrument
Bronkhorst High-Tech B.V.IQ+FLOW 9.17.045102.10 ZeroingThe zero point of each instrument is factory adjusted. If required the zero point can be re-adj
Bronkhorst High-Tech B.V.IQ+FLOW 119.17.0452.13 Model key2.13.1 Model key IQF / IQP2.13.2 Model key IQM* Each channel of an instrument will be provide
Bronkhorst High-Tech B.V.IQ+FLOW 9.17.045123 Basic operation3.1 GeneralAn IQ+FLOW® instrument must be powered with +15…+24 Vdc according to the applic
Bronkhorst High-Tech B.V.IQ+FLOW 139.17.0453.3 Manual interface: micro-switch, LEDs and rotary switchesIn this section the manual instrument interface
Bronkhorst High-Tech B.V.IQ+FLOW 9.17.04514LEDsTimePushedIndicationGreen Red OffOff0 – 4 sec.No action. Pressing a switch shortly by accident will not
Bronkhorst High-Tech B.V.IQ+FLOW 159.17.0453.4Basic RS232 operationDigital operation adds a lot of extra features (compared to analog operation) to th
Bronkhorst High-Tech B.V.IQ+FLOW 9.17.045163.4.4SoftwareExamples of free Bronkhorst® DDE client applications: FlowDDE, FlowPlot and FlowView. Other so
Bronkhorst High-Tech B.V.IQ+FLOW 179.17.0453.5Basic RS485 operationThis section is limited to the description of the interface between the IQ+FLOW® in
Bronkhorst High-Tech B.V.IQ+FLOW 9.17.045183.5.2SoftwareWhen using a pc to communicate with IQ+FLOW® instruments only the FLOW-BUS protocol is support
Bronkhorst High-Tech B.V.IQ+FLOW 199.17.0453.6Basic parameters and properties3.6.1 IntroductionMost instrument parameters can only be accessed with di
DisclaimerThe information in this manual has been reviewed and is believed to be wholly reliable. No responsibility, however, is assumed forinaccuraci
Bronkhorst High-Tech B.V.IQ+FLOW 9.17.045203.6.2Basic measurement and control parametersThe list below provides the most basic parameters for digital
Bronkhorst High-Tech B.V.IQ+FLOW 219.17.0454 Advanced operation4.1 Reading and changing instrument parameters4.1.1 Special parametersAll parameters de
Bronkhorst High-Tech B.V.IQ+FLOW 9.17.0452220Valve steeringSetpoint is redirected to Valve Out withcontroller idle21Analog valvesteeringAnalog input i
Bronkhorst High-Tech B.V.IQ+FLOW 239.17.0454.1.3 Advanced measurement and control parametersMeasured Value (Fmeasure)TypeAccessRangeFlowDDEFLOW-BUSMod
Bronkhorst High-Tech B.V.IQ+FLOW 9.17.04524Controller Speed (Kspeed)TypeAccessRangeFlowDDEFLOW-BUSModbusFloatRW 0…3.4E+38254114/300xF2F0…0xF2F1/62193…
Bronkhorst High-Tech B.V.IQ+FLOW 259.17.045Dynamic Display Factor TypeAccessRangeFlowDDEFLOW-BUSModbusFloatRW 0…1.056117/10xF508…0xF509/62729…62730Thi
Bronkhorst High-Tech B.V.IQ+FLOW 9.17.04526Auto-zero procedure:Step 1: Set 'Control Mode' to 'Calibration Mode' (value 9)Step 2: S
Bronkhorst High-Tech B.V.IQ+FLOW 279.17.0454.2.4 Setting digital output (multi-channel versions only)The IQ+FLOW® multi-channel pc-board is equipped w
Bronkhorst High-Tech B.V.IQ+FLOW 9.17.04528Value zero will be indicated by a period of 1 sec off (0.5 sec off + 0,5 sec off).When value zero is wanted
Bronkhorst High-Tech B.V.IQ+FLOW 299.17.0455 Troubleshooting and service5.1 GeneralFor a correct analysis of the proper operation of an instrument it
WarrantyBronkhorst products are warranted against defects in material and workmanship for a period of three years from the date ofshipment provided th
Bronkhorst High-Tech B.V.IQ+FLOW 9.17.045305.4ServiceFor current information on Bronkhorst High-Tech B.V. and service addresses please visit our websi
Bronkhorst High-Tech B.V.IQ+FLOW 9.17.0454Table of contents...
Bronkhorst High-Tech B.V.IQ+FLOW 59.17.045...
Bronkhorst High-Tech B.V.IQ+FLOW 9.17.04561 Scope of this manual1.1 IntroductionThis manual covers the IQ+FLOW® series mass flow/pressure meters/contr
Bronkhorst High-Tech B.V.IQ+FLOW 79.17.0452 Starting up2.1 Check properties1. Check IQ+FLOW® propertiesBefore installing it is important to read the a
Bronkhorst High-Tech B.V.IQ+FLOW 9.17.0458Compression type fittingsFor leak tight installation of compression type fittings make sure that the tube is
Bronkhorst High-Tech B.V.IQ+FLOW 99.17.045Digital RS232 operationDigital operation over RS232 can be establishedwhen using the following setup. Using