Bronkhorst High-Tech B.V.
IQ+FLOW 21
9.17.045
4 Advanced operation
4.1 Reading and changing instrument parameters
4.1.1 Special parameters
All parameters described in this chapter have influence on the behaviour of the IQ
+
FLOW®. Please be aware that
wrong settings can disorder the output. To avoid unintential changes, some parameters are locked (shown by the
symbol). To unlock parameters set parameter 'Init Reset' to 'Unlocked'.
Init Reset
The 'Init Reset' parameter is used to unlock advanced parameters for writing. This parameter can be set to the following values:
Advanced parameters are read-only
Advanced parameters are writeable and readable
This parameter is always set to 'Locked' at power-up.
Control Mode
The 'Control mode' is used to select different functions of the instrument. The following modes are available:
Controlling as slave of other instrument on bus
FLOW-BUS *
slave factor /
100%
Stand-by on BUS/RS232, Controlling is stopped;
Valve Out freezes in current position
Test mode enabled (factory only)
Tuning mode enabled (factory only)
Controlling at setpoint 100%
Calibration mode enabled (factory only)
Controlling as slave of other instrument on
analog input
Analog input *
slave factor /
100%
Controlling at setpoint 0%
Controlling as slave of other instrument on bus,
slave factor is set with signal on analog input
FLOW-BUS *
analog input *
slave factor /
100%