Page 44 Operational instructions for digital multibus instruments 9.17.023
11.2.3 LED indications using micro-switch at normal running mode of an instrument
When the switch is pressed-down both LED’s will be switched-off for function selection. As long as the switch will be
pressed-down, there will be a change in indication by the 2 LED’s after each 4 seconds. The moment the user
recognizes the indication (LED-pattern) for the function he wants, he must release the switch. Now the wanted
function is triggered.
Pressing a switch shortly by accident will not cause unwanted reactions of instrument.
In case of min/max alarm or counter batch reached:
Reset alarm (only if reset by keyboard has been enabled)
See specific field bus for more details.
When address is occupied:
Automatic installation on FLOW-BUS.
Instrument program will be restarted and all warning and error message will be cleared.
During a start-up the instrument will perform a self-test
Instrument will be re-adjusted for measurement of zero-flow (not for pressure
meter/controller)
NOTE: First make sure there is no flow and instrument is connected to power for at
Set instrument in the FLASH mode
This mode will be indicated by both LED’s off when instrument is normally powered
11.2.4 LED indications using micro-switch at power-up situation
Here is described what the indications are for the functions to be performed at power-up situation of an instrument.
This can be realized by pressing the switch first and while pressing, connecting the power. These actions have a more
‘initializing’ character for the instrument.
Pressing a switch shortly by accident will not cause unwanted reactions of the
0.2 sec on,
All parameter settings (except field bus settings) will be restored to situation of
final test at BHT production.
0.2 sec on,
0.2 sec off
See specific field bus for more details.
Auto install to bus Instrument will install itself
to a (new) free node-address on the FLOW-BUS.
0.2 sec on,
0.2 sec off
0.2 sec on,
0.2 sec off
For MBC-II type of instruments, the default address will be set immediately.
The default address will be set after leaving this mode (approx. 60 sec)
See specific field bus for default installation address: