BRONKHORST HIGH-TECH B.V.
2.13.7 Needle Valves
- Removal
Ensurethatthesystemisdepressurisedbeforetakingthe
needlevalveapart.
Loosenthe4boltswithhexagonheadcounterclockwise
toreleasetheneedlevalve.
9.17.057 page 14
- Replacement
ReplacethegasketsAS‐007andAS‐017,Viton,
EPDMorKalrez.
Allwettedpartshavetobeabsolutelyclean,don’tuseoil
orgreaseandavoiddust!
Replacetheneedlevalvecentrebeforetightenthebolt
withhexagonheadbyturningclockwise,tightencrosswise
withM30,9Nm.
Nowtheneedlevalvehastobetestedforleakageandfunction.
- Adjustment
Theneedlevalvesareadjustable
withaslottedheadscrewdrivertoadjusttheflow.
TheNV‐01isalsolockablewitha
nutbecausethefineadjustability.
AvailableNeedlevalves
Modelnumber Description
NV‐01 Kvmin1.0x10
‐3
,Kvmax6.8x10
‐3
Lowflowneedlevalve,fine
NV‐02 Kvmin5.6x10
‐3
,Kvmax5.6x10
‐2
Highflowneedlevalve,coarse
2.13.8 Pneumatic operated shut off valves NC or NO
Theactuatingpressurehastobesuppliedwitha3/2wayvalve.
Whenthevalveisnotactuatedtheairactuationinputhastovent.
Actuationpressure :
Minimal :3 bar(g).
Maximal :9 bar(g).
- Removing
Ensurethatthesystemisdepressurisedbeforetaking
thepneumaticshutoffvalveapart.
Removetheactuatingpressuretubingandloosenthe4bolts
withhexagonheadcounterclockwisetoreleasethevalve.
- Replacing
ReplacethegasketAS‐007andAS‐017,Viton,EPDMorKalrez.
Allwettedpartshavetobeabsolutelyclean,don’tuse
oilorgreaseandavoiddust!
Replacethepneumaticshutoffvalveandtightenthe
boltswithhexagonheadbyturningclockwise,tighten
crosswisewithM30,9Nm.
Replacetheactuatingpressuretubing.
Nowthepneumaticshutoffvalvehastobetestedforleakageandfunction.
AvailablePneumaticshutoffvalves
Modelnumber Description
PV‐01‐NC‐V Kv=0,20
NormallyClosedViton
PV‐01‐NC‐E Kv=0,20
NormallyClosedEPDM
PV‐01‐NC‐K Kv=0,20
NormallyClosedKalrez
PV‐01‐NO‐V Kv=0,20
NormallyOpenViton
PV‐01‐NO‐E Kv=0,20
NormallyOpenEPDM
PV‐01‐NO‐K Kv=0,20
NormallyOpenKalrez